没有合适的资源?快使用搜索试试~ 我知道了~
Half-Fanbeam Collimators Combined with Scanning Point Sources fo...
需积分: 9 5 下载量 143 浏览量
2010-03-24
17:13:37
上传
评论
收藏 1.86MB PDF 举报
温馨提示
试读
8页
这是从国外网站下载的关于CT重建的文献,这个资源就不扣大家分数了,大家下载东西都不容易,积分也不容易,希望大家多上传资料共享
资源推荐
资源详情
资源评论
Half-Fanbeam Collimators Combined with Scanning
Point Sources for Simultaneous Emission-
Transmission Imaging
Freek J. Beekman, Chris Kamphuis, Brian F. Hutton and Peter P. van Rijk
Department of Nuclear Medicine, Image Sciences Institute, University Hospital Utrecht, Utrecht, The Netherlands; and
Department of Medical Physics, Westmead Hospital, Sydney, Australia
One type of SPECT system often used for simultaneous emission-
transmission tomography is equipped with parallel-hole collimators,
moving line sources (MLS) and electronic windows that move in
synchrony with the sources. Although downscatter from the emis
sion distribution is reduced by the use of the electronic window, this
still can represent a sizable fraction of the transmitted counts. These
systems have relatively poor spatial resolution and use costly
transmission sources. Methods: Using a two-head SPECT system,
with heads at right angles, two 153Gdline sources (5800 MBq each)
were replaced by two 153Gd point sources of only 750 MBq each
and positioned to move along the focal lines of two half-fanbeam
collimators. A suitable acquisition protocol for a moving point source
(MPS) system was selected by considering the results of a simula
tion study. With this protocol, physical phantom experiments were
conducted. Results: Simulations showed that by using two half-
fanbeam collimators, a gantry rotation of 90°,such as used for 180°
acquisition with parallel-beam collimators for cardiac imaging, was
insufficient. A gantry rotation of 180°resulted in attenuation maps
where only an area to the posterior of a 400-mm wide thorax
phantom was affected by truncation. The MPS system had a 14.7
times higher sensitivity for transmission counts than the MLS sys
tem. Despite the smaller sources in the MPS system, the number of
acquired transmission counts was a factor 1.91 times higher com
pared with the MLS system, resulting in reduced noise. The relative
downscatter contribution from 99mTc(140 keV) in the 153Gdmoving
electronic window (100 keV) was reduced by a factor of 1.81.
Transmission images of a rod phantom with segments containing
acrylic rods of different diameters showed an improvement of
resolution in favor of the MPS system from about 11 mm to about 6
mm (five instead of two segments of rods were clearly visible). In
addition, the noise level in the MPS thorax transmission images was
significantly lower. Conclusion: The MPS system has important
advantages when compared with the MLS system. The use of
low-activity point sources is economically beneficial when com
pared with line sources and reduces radiation exposure to staff and
patients.
Key Words: quantitative SPECT; attenuation correction;transmis
sion scanning; fanbeam collimator
J NucÃ-Med 1998; 39:1996-2003
Phcbotónattenuation in patients greatly degrades quantitation of
SPECT images and introduces image artifacts and distortions.
Accurate correction for attenuation can be performed when the
density distribution of the patient is known [for example from
transmission CT (TCT) images]. Attenuation correction using
separate transmission CT is difficult, since the three-dimen
sional radiograph image needs to be registered accurately to the
SPECT scan. Much effort has been expended during the last
decade in developing SPECT systems that acquire TCT data
simultaneously with the emission data [emission CT (ECT)-
Received Aug. 5, 1997; revistan accepted Feb. 18, 1998.
For correspondence or reprints contact: Freek J. Beekman PhD, University Hospital,
Utrecht, E 02.222, Heidelberglaan 100, 3584 CX Utrecht. The Netherlands.
TCT scanning]. The use of such systems can improve signifi
cantly the diagnostic accuracy of cardiac SPECT for the
detection and localization of coronary heart disease (7). In
addition to attenuation correction, transmission maps can be
useful for anatomical localization of activity (e.g., in tumors and
infectious foci), for registration of images from other imaging
modalities, for dose calculations (2-4) and attenuation, map-
based, scatter correction (5-11). A large variety of hardware
configurations and reconstruction algorithms for combined
ECT-TCT imaging are summarized in (12,13).
One class of ECT-TCT systems uses parallel-hole collimators
combined with sheet sources (14-19) (Figure 1A). To reduce
both radiation dose to the patient and scatter and to improve
resolution, the sheet source may be collimated (14,20). ECT-
TCT systems equipped with parallel-hole collimators have been
improved significantly by using high-intensity, moving line
sources and an electronic window moving in synchrony with
the source ["electronic collimation" (21)} (Fig. IB). The mov
ing electronic window defines a "strip" on the detector in which
transmission counts are collected. Counts on the remainder of
the detector area are collected into the emission image. In the
case of a parallel-hole collimator and a flood source (Fig. 1A),
a low number of transmission counts is obtained and a high
amount of downscatter from the emission distribution is de
tected. When the transmission source is a concentrated moving
line source, the same amount of transmission counts are
acquired in the moving electronic window (Fig. IB), but the
downscatter contribution is lower since downscatter now is
acquired only during a fraction of the total acquisition time, as
defined by the ratio of the electronic window width to the field
size. Therefore, the moving line source (MLS) system permits
improved separation of ECT and TCT data and possible use of
transmission sources with a lower energy than the radiophar-
maceutical, with significantly reduced downscatter, as is shown
by the graphs in the bottom row of Figure 1. An MLS system,
with two camera heads at right angles, is shown at the left in
Figure 2. An advantage of parallel-hole geometries, compared
with most converging geometries, is that truncation of the
subject in projection images is avoided in nearly all cases
provided the camera field size is sufficiently large. The disad
vantage is that the required activity of a fresh 153Gdline source
is relatively high (typically 100-250 mCi). Even with this
source activity, very few TCT counts are acquired in some
pixels when large patients are scanned. This leads to problems
in reconstruction, especially when these low-count pixels have
to be corrected for downscatter from 99mTc. Regular replace
ment of 153Gd line sources (half-life 241.6 days) is expensive,
and difficulties have been experienced by some departments in
obtaining licenses for the use of these potent sources. The
degrading effects of downscatter can be corrected (21,22),
albeit at the cost of some noise amplification. This noise
1996 THEJOURNALOFNUCLEARMEDICINE•Vol. 39 •No. 11 •November 1998
资源评论
shamorenhai
- 粉丝: 0
- 资源: 4
上传资源 快速赚钱
- 我的内容管理 展开
- 我的资源 快来上传第一个资源
- 我的收益 登录查看自己的收益
- 我的积分 登录查看自己的积分
- 我的C币 登录后查看C币余额
- 我的收藏
- 我的下载
- 下载帮助
安全验证
文档复制为VIP权益,开通VIP直接复制
信息提交成功