X-StreamDSO Firmware Update
The firmware installer updates multiple oscilloscope
components including the base application, required
DLLs, device drivers and low-level microcode for
integrated circuits.
The update does not modify or delete any saved panel
setups, waveforms, screen captures, calibration
constants, or other data stored on the D: drive.
Caution: Do not install any XStreamDSO version
prior to 8.6.1.x on Windows 10 oscilloscopes,
and 8.6.1.8 on WavePro HD oscilloscopes in
particular. Doing so will prevent the oscilloscope
from running in the non-administrative user
account, and you will need to run the recovery
procedure. To install firmware or third-party
software on Windows 10 machines, you must
be logged on as user LCRYADMIN or supply the
password SCOPEADMIN.The installation may
take several minutes. Do not power down at any
point during the process.
Download Instructions
Tip: If the oscilloscope has an internet
connection, choose File > Exit to close
X-StreamDSO and use the oscilloscope
browser to download the installer directly.
1. Visit our download page at
teledynelecroy.com/support/softwaredownload and
go to Oscilloscope Downloads > Firmware Upgrades.
2. Select your series and model number.
3. Enter your registration login information, or create a
new account. You cannot proceed without an account.
4. Click the download link, and choose to Save the
installer to the oscilloscope Desktop or to a USB
storage device for transfer to the instrument.
Installation Instructions
1. From the oscilloscope menu bar, choose File > Exit to
close X-StreamDSO.
2. Browse to xstreamdsoinstaller_x.x.x.x.exe and
double-click it to launch the setup wizard.
3. On Windows 10 instruments running in Non-Admin
mode (the default state), enter the Password
SCOPEADMIN (all uppercase).
4. Follow the installer prompts. Leave the default
installation (recommended), or select individual
components:
l Drivers for GPIB1 - required for GPIB card
l MATLAB MCR - required for sampling scopes,
QPHY-USB and QPHY-BroadR-Reach
l X-Stream DSO, DSO Device drivers, Upgrade DSO
Microcode - required for version upgrade
l Drivers for CAN - required for external CAN trigger
module
l Touch Screen Driver - required to use the display
as a touch screen
l Drivers for MSxx - required for Mixed Signal
devices MS-32, MS-250 and MS-500
l LSIB Package - required for LSIB
hostboard/hostcard
l SPARQ Package - required to drive a connected
SPARQ from the instrument
Click Install when done.
Note: Not all components are available on all
models. If you do not see an option above
listed in your installer, it is not required for
your instrument.
5. If you receive Windows security warnings, trust and
Install the file.
6. If you see the Hardware Programmers screen,
accept all code installations, then click the closebox
after you see that programming has completed to
return to the X-Stream Setup wizard.
7. When installation is complete, choose Reboot now
and click Finish.
Release Notes
Release Notes are provided for each new version. They
can be found on the series download page where you
downloaded the installer.
Note to QualiPHY Users
X-StreamDSO and QualiPHY software versions must
match to a point-release to function properly. If you are
upgrading your oscilloscope firmware, upgrade your
version of QualiPHY as soon as possible.
© 2018 Teledyne LeCroy, Inc. All rights reserved.
923614 Rev E
Installation Instructions