Equipment for neonatal and premature infants: Incubator
Atom Infant Incubator 100
Operation Manual
TO THE OPERATOR AND THE PERSON IN CHARGE OF
MAINTENANCE AND CARE OF THE UNIT:
Read this ● Manual carefully before operating the unit.
Keep this ● Manual where it is readily accessible for reference when needed.
This Manual contains instructions for operating the oxygen controller, the ●
weight monitor and the pulse oximeter. Please skip any section unrelated to
your unit.
ORIGINAL
Via Libia, 54 - 20081 Abbiategrasso (MI) - Italy
Tel: +39 02 99763101 Fax: +39 02 99763110
EU OFFICE
1
INTRODUCTION
This Operation Manual deals with the specifications, operation and maintenance of the Dual Incu
i
. Atom is by no
means responsible for any malfunction arising from a user ignoring the instructions for operation and maintenance
described in this Manual as well as any accident attributable to repair by someone other than technical personnel
belonging to or authorized by Atom.
This Manual contains instructions for operating the oxygen controller, the weight monitor and the pulse oximeter.
Please skip any section unrelated to your unit.
Read this Manual carefully and familiarize yourself with its contents before operating the unit. Keep this Manual
where it is readily accessible for reference when needed. If any technical problem should arise, please contact
your local Atom representative.
This product is shipped without being disinfected. Be sure to clean and disinfect the unit before us-
ing it for the first time after purchase.
This product maintains the ambient temperature around an infant by drawing in the air outside and
warming it. It does not have the capability to cool the air drawn from outside. Be sure to set the
incubator air temperature at least 3°C higher than the ambient temperature. In particular, if a photo-
therapy unit, a heated humidifier, or some other heat-generating device is used with the incubator,
set the incubator air temperature at least 5°C higher than the ambient temperature. If a lower setting
is selected, the incubator may not be controlled correctly.
INTENDED USE
The Dual Incu
i
is an incubator for newborns and premature neonates. It is mainly intended for temperature control
in neonatal hypothermia, observation and examination in newborn nurseries, prevention of body temperature drop
shortly after delivery, and pre-operative and post-operative intensive care in neonatal surgery. This incubator has
the capability to control the infant’s skin temperature.
SAFETY INFORMATION
Instructions to ensure the safe operation of the unit are found throughout this Manual. Please read the Manual
carefully before operating the unit. Please follow the instructions when operating the unit.
Basic Instructions
1. Follow the instructions for the safe use of the unit.
Follow the operating instructions described in this Manual for the safe use of the unit.
2. Inspect the unit on a periodical basis.
Periodical inspection is needed to use the unit in the optimum condition.
3. Never use the unit if it is found to be defective.
If any damage or malfunction of the unit should be noticed, stop using it immediately and contact your local
Atom representative.
4. Follow the EMC information given in this Manual.
Electric equipment for medical use needs special precautions regarding EMC. It needs to be installed and put
into service according to the EMC information provided in this Manual.
2
Definitions of Warning Indication
Three levels of warning indication are used throughout this Manual and on the unit. They are defined as
follows.
A DANGER notice indicates an immediately hazardous situation which, if not avoided,
will result in death or serious injury, serious damage to property such as total loss of use
of equipment, or fire.
A WARNING notice indicates an indirectly (potentially) hazardous situation which,
if not avoided, will result in death or serious injury, serious damage to property such as
total loss of use of equipment, or fire.
A CAUTION notice indicates a hazardous situation which, if not avoided, can result in
minor or moderate injury, partial damage to equipment, and loss of data stored
in computers.
Definition of Symbols
1. Symbols to indicate danger, warning or caution
Symbol Title and indication
General attention
Indicates unspecified general danger, warning or caution.
Caution: Hot surface
Indicates that the surface can be dangerously hot under certain conditions.
2. Symbols to prohibit action
Symbol Title and indication
General prohibition
Indicates unspecified general prohibition.
Prohibition of disassembly
Indicates prohibition of disassembly of the unit where it may cause an electric shock or other
hazards.
Prohibition of use of fire
Indicates prohibition of use of fire where an external use of fire may cause the unit to ignite under
certain conditions.
3. Symbols to give instructions for action
Symbol Title and indication
General instruction
Indicates unspecified general action on the part of the user.
Connect a ground wire
Instructs the user to connect the ground wire without fail where the unit is provided with a ground
terminal.
Remove the power plug from the power outlet
Instructs the user to remove the power plug from the power outlet in the case of malfunction or
when there is a threat of lightning.
3
4. Symbols of international standards (IEC)
Symbol Title and indication
Power on
Indicates in relation to the power switch that the device is connected to the power source.
Power off
Indicates in relation to the power switch that the device is not connected to the power source.
Equipment partially on
Indicates that a part of the device is “ON.”
Equipment partially off
Indicates that a part of the device is “OFF.”
Type BF applied part
Indicates that the device is classified as Type BF in terms of the degree of protection against an
electric shock.
Bell silenced
Indicates either a control switch to silence the bell permanently or temporarily, or that the bell is
silenced.
Manufacturer
This symbol indicates the name and the address shown adjacent to the symbol is of the manufac-
turer.
Date of manufacture
Indicates the date when the unit was manufactured in the factory.
Authorised representative in the european community
This symbol indicates the name and the address shown adjacent to the symbol is of the authorised
representative in the European Community.
WEEE symbol
In the EC area, an electrical and electronic product falling in one of the categories specified by
“DIRECTIVE 2002/96/EC OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 27
January 2003 on Waste Electrical and Electronic Equipment (WEEE)” should be disposed of in a
manner consistent with relevant laws and regulations.
This symbol indicates that the above-mentioned requirement applies to this product.
Locked
Indicates a locked condition.
Unlocked
Indicates an unlocked condition.
See Operation Manual
Follow Operation Manual.
剩余147页未读,继续阅读
资源评论
qq_30216093
- 粉丝: 0
- 资源: 1
上传资源 快速赚钱
- 我的内容管理 展开
- 我的资源 快来上传第一个资源
- 我的收益 登录查看自己的收益
- 我的积分 登录查看自己的积分
- 我的C币 登录后查看C币余额
- 我的收藏
- 我的下载
- 下载帮助
最新资源
- Wireshark-4.2.4-x64.zip
- 2022年11月软件设计师上
- 基于VB+ACCESS教学管理系统(参考文献+源代码).zip
- 一个工具的流程图 demo
- EMC3080的用于连接FogCloud的固件
- 基于VB+Access酒店客房管理系统(源代码+参考文献+报告).zip
- 应用笔记LAT1244+奇怪的NRST+管脚异常复位问题
- SEMI标准的解释说明
- tensorflow-2.8.2-cp37-cp37m-manylinux2010-x86-64.whl
- tensorflow-rocm-2.12.0.560-cp311-cp311-manylinux2014-x86-64.whl
资源上传下载、课程学习等过程中有任何疑问或建议,欢迎提出宝贵意见哦~我们会及时处理!
点击此处反馈
安全验证
文档复制为VIP权益,开通VIP直接复制
信息提交成功